DIFFERENTIAL PROBE 9322

Description

3 Kinds of Measurements with a Single Probe

Basic specifications

Accuracy guaranteed: 1 year
*1: Discontinued product

Measurement functions DC mode: Waveform monitor output, DC to 10 MHz ±3 dB
AC mode: Detection of power line surge noise, 1 kHz to 10 MHz ±3 dB
(Low frequency cut-off frequency 1 kHz ± 300 Hz)
RMS mode: Rectified RMS output of DC and AC voltages, DC, 40 Hz to 100 kHz, Response speed: 200 ms or less (400 V AC)
Max. allowable input 2000 V DC, 1000 V AC
Max. rated voltage to earth When using the Grabber Clip L9243: 1000 V AC/DC (CAT II)
When using alligator clip: 1000 V AC/DC (CAT II), 600 V AC/DC (CAT III)
Output Voltage division ratio: 1/1000, BNC terminal (DC/AC/RMS 3-mode selectable output)
DC amplitude accuracy ±1 % f.s. (1000 V DC or less), ±3 % f.s. (2000 V DC or less) (f.s.=2000 V DC)
RMS amplitude accuracy ±1 % f.s. (DC, 40 Hz to 1 kHz), ±4 % f.s. (1 kHz to 100 kHz) (f.s.=1000 VAC)
Input resistance, capacity H-L: 9 MΩ, approx 10 pF (C at 100 kHz)
H-case, L-case: 4.5 MΩ, approx 20 pF (C at 100 kHz)
Power supply +5 V to +12 V, less than 300 mA (DC jack OD 5.5 mm [0.22 in.], ID 2.1 mm [0.08 in.])

- Via AC Adapter 9418-15
- Via MR6000 dedicated Probe Power Unit Z5021 through Power Cord 9248
- Via Logic terminal on Memory HiCorder through Power Cord 9324 (*1)
- Via sensor terminal of F/V Unit 8940 (*1) through Power Cord 9325 (*1)
- Via DC power output terminal attached to the input unit for the 8855 through Power Cord 9328 (*1)
- Via the 8860 series dedicated Probe Power Unit 9687 (*1) through Power Cord 9248
Dimensions and mass 70 mm (2.76 in)W × 150 mm (5.91 in)H × 25 mm (0.98 in)D, 350 g (12.3 oz), Cord length: Input 46 cm (1.51 ft), Output 1.3 m (4.27 ft)
Included accessories Alligator clips ×1 (red/black set), Grabber Clip L9243 ×1 (red/black set), Carrying case C0203 ×1, Instruction manual ×1