3 Kinds of Measurements with a Single Probe
Accuracy guaranteed: 1 year
*1: Discontinued product
Measurement functions | DC mode: Waveform monitor output, DC to 10 MHz ±3 dB AC mode: Detection of power line surge noise, 1 kHz to 10 MHz ±3 dB (Low frequency cut-off frequency 1 kHz ± 300 Hz) RMS mode: Rectified RMS output of DC and AC voltages, DC, 40 Hz to 100 kHz, Response speed: 200 ms or less (400 V AC) |
|||||
---|---|---|---|---|---|---|
Max. allowable input | 2000 V DC, 1000 V AC | |||||
Max. rated voltage to earth | When using the Grabber Clip L9243: 1000 V AC/DC (CAT II) When using alligator clip: 1000 V AC/DC (CAT II), 600 V AC/DC (CAT III) |
|||||
Output | Voltage division ratio: 1/1000, BNC terminal (DC/AC/RMS 3-mode selectable output) | |||||
DC amplitude accuracy | ±1 % f.s. (1000 V DC or less), ±3 % f.s. (2000 V DC or less) (f.s.=2000 V DC) | |||||
RMS amplitude accuracy | ±1 % f.s. (DC, 40 Hz to 1 kHz), ±4 % f.s. (1 kHz to 100 kHz) (f.s.=1000 VAC) | |||||
Input resistance, capacity | H-L: 9 MΩ, approx 10 pF (C at 100 kHz) H-case, L-case: 4.5 MΩ, approx 20 pF (C at 100 kHz) |
|||||
Power supply | +5 V to +12 V, less than 300 mA (DC jack OD 5.5 mm [0.22 in.], ID 2.1 mm [0.08 in.]) - Via AC Adapter 9418-15 - Via MR6000 dedicated Probe Power Unit Z5021 through Power Cord 9248 - Via Logic terminal on Memory HiCorder through Power Cord 9324 (*1) - Via sensor terminal of F/V Unit 8940 (*1) through Power Cord 9325 (*1) - Via DC power output terminal attached to the input unit for the 8855 through Power Cord 9328 (*1) - Via the 8860 series dedicated Probe Power Unit 9687 (*1) through Power Cord 9248 |
|||||
Dimensions and mass | 70 mm (2.76 in)W × 150 mm (5.91 in)H × 25 mm (0.98 in)D, 350 g (12.3 oz), Cord length: Input 46 cm (1.51 ft), Output 1.3 m (4.27 ft) | |||||
Included accessories | Alligator clips ×1 (red/black set), Grabber Clip L9243 ×1 (red/black set), Carrying case C0203 ×1, Instruction manual ×1 |