AC/DC CURRENT PROBE CT6846A

Description

High-precision Current Testing

Basic specifications

Accuracy guaranteed: 1 year

Rated current 1000 A AC/DC
Frequency bandwidth DC to 100 kHz
Diameter of measurable conductors Max. φ 50 mm (1.97 in.)
Max. allowable input ±1900 Apeak
(Within 20 ms in an environment of 40°C/104°F or less)
Output voltage 2 mV/A
Output resistance 50 Ω ± 10 Ω
Accuracy (amplitude)
±(% of reading + % of full scale)
DC: 0.2% + 0.02%
DC < f ≤ 100 Hz: 0.2% + 0.01%
Linearity ±20 ppm Typical
Common-Mode Voltage Rejection Ratio
(CMRR)
DC to 1 kHz:150 dB or greater
1 kHz to 10kHz:130 dB or greater
10 kHz to 50 kHz:100 dB or greater
(effect on output voltage and common mode voltage)
Automatic phase correction Automatically performs phase correction when connected to PW8001
Operating temperature, humidity -40°C to 85°C (-40°F to 185°F), 80% RH or less
Standards Safety IEC 61010-2-032:2012/EN 61010-2-032:2012 Type D
EMC IEC 61326-1:2012/EN 61326-1:2013
Withstand voltage AC 4,260 V
Power supply Power suppled via the Power Analyzer PW8001, PW6001, PW3390, Sensor Unit CT9555, CT9556, CT9557, or 3CH CURRENT UNIT U8977
Max. rated power 7 VA or less (When measuring 1000 A/55 Hz with ±12 V power supply)
Dimensions and mass Approx. 238 mm W × 116 mm H × 35 mm D
(approx. 9.37 in. W × 4.57 in. H × 1.38 in. D), 990 g
Cable length Approx. 3 m (9.84 ft.)
Included accessories ・Markup bands(×6)
・Carrying case
・Instruction manual