Best-in-class Measurement Bandwidth with High Accuracy
Accuracy guaranteed: 1 year
Rated current | 800 A AC/DC |
---|---|
Max. allowable input | Within the derating range, However, up to ±1200 A peak is allowable if within 20 ms. (design value) |
Frequency bandwidth | Amplitude: DC to 4 MHz (CT6904A-2), DC to 2 MHz (CT6904A-3) Phase: DC to 1 MHz |
Linearity | ±12.5 ppm Typical (23°C (73°F)) |
Offset voltage | ±10 ppm Typical (23°C (73°F), no input) |
Basic accuracy | DC (Amplitude: ±0.030 % rdg. ±0.009 % f.s., No phase specification) 45 Hz ≤ f ≤ 65 Hz (Amplitude: ±0.025 % rdg. ±0.009 % f.s., Phase: ±0.08°) |
Output voltage rate | 2 mV / A rated (This device outputs AC+DC voltage via the Sensor Unit) |
Max. rated voltage to earth | 1000 V CAT III |
Core diameter | φ 32 mm (1.26 in) |
Operating temperature, humidity | -10°C to 50°C (14°F to 122°F) 80% RH or less (with no condensation) |
Power supply | Power suppled via the Power Analyzer PW8001, PW6001, PW3390, or Sensor Unit CT9555, CT9556, CT9557 |
Max. rated power | 7 VA Max. (800 A/55 Hz measurement, with a power supply of ±12 V) |
Dimensions and mass | 139 mm (5.47 in)W × 120 mm (4.72 in)H × 52 mm (2.05 in)D, CT6904A-2: 1.15 kg (40.6 oz), cable length 3 m (9.84 ft), CT6904A-3: 1.45 kg (51.1 oz), cable length 10 m (32.81 ft) |
Included accessories | Instruction Manual ×1, Carrying case ×1, Color labels (for channel identification) ×1, Operating Precautions ×1 |