AC/DC CURRENT SENSOR CT6872

Description

Low-current Model of 50 A rating, with Wideband and High Accuracy

Basic specifications

Accuracy guaranteed: 1 year

Rated current 50 A AC/DC
Max. allowable input Within the derating range, up to ±150 A peak (design value) allowed at 40°C or less for 20 ms or less
Frequency bandwidth Amplitude: DC to 10 MHz
Phase: DC to 1 MHz
Basic accuracy DC (±0.03% rdg. ±0.002% f.s., no phase specification)
45 Hz ≤ f ≤ 66 Hz (±0.03% rdg. ±0.007% f.s., ±0.05°)
Specified up to 1 MHz
Output voltage rate 40 mV/A rated (This device outputs AC+DC voltage via the Sensor Unit)
Max. rated voltage to earth 1000 V CAT III
Core diameter φ 24 mm (0.94 in)
Operating temperature, humidity -40°C to +85°C (-40°F to 185°F), 80% RH or less (with no condensation)
Power supply Power suppled via the Power Analyzer PW8001, PW6001, PW3390, Sensor Unit CT9555, CT9556, CT9557, or 3CH CURRENT UNIT U8977
Max. rated power 4 VA max. (at 50 A/55 Hz, ±12 V power requirement)
Dimensions and mass 70 mm (2.76 in)W × 100 mm (3.94 in)H × 53 mm (2.09 in)D, CT6872: 370 g (13.1 oz), cable length: 3 m (9.84 ft), CT6872-01: 690g (24.3 oz), cable length 10 m (32.81 ft)
Included accessories Instruction Manual ×1, Mark bands ×6, Operating Precautions ×1

Compatible models

Compatible models CT6872
Power Analyzer
PW8001, PW6001, PW3390
OK
Power Analyzer
3390
Attention (Requires CT9901)
Power HiTester
3193 series
Attention (Requires CT9901)
Current Unit
8971
Attention (Requires the 9318, CT9901)
3CH Current Unit U8977 OK